Heath Consultants Remote Methane Leak Detector Intrinsically Safe (RMLD-I) User Manual
Page 19

• BACLI
i. Press up button to turn on or to turn off the display backlight.
Fig. 3-5
 
Although the ppm-m readout will not be updating, the instrument will still operate 
properly while in a menu option other than the run display. 
Using the DMD Mode 
 
The Digital Methane Detection (DMD) mode is a highly sophisticated detection 
algorithm that greatly enhances the use of the RMLD-IS. In most situations, the operator 
should survey with the DMD mode turned on. To turn on the DMD mode, press the 
DMD button (the DMD icon will appear on the display). This mode can only be activated 
while in the main survey screen. 
 
Fig. 3-6: DMD icon is displayed when in DMD mode
 
While using the DMD, no sound will be heard until a detection of methane occurs. The 
pitch of the sound will be relative to the methane concentration. The higher the pitch, the 
stronger the methane concentration is. 
 
A low-pulsating or continuous sound will indicate a warning due to an infrared laser low 
light level condition or instrument fault. The warning icon will also be displayed 
indicating that the light level is too low (See Long Range Scanning Pg. 24). The operator 
needs to move in closer to get in range. 
 
Should the warning persist, it could be due to a fault in the instrument. Check the error 
code being displayed and follow the instructions in the troubleshooting guideline. 
 
DMD icon is on
101296-0 RMLD IS MANUAL REV J