Hvc troubleshooting – Heath Consultants Heath Volume Corrector (HVC) User Manual
Page 26
26
HVC Troubleshooting
The following may serve as a quick troubleshooting guide if you encounter problems in operating the
HVC:
Symptom
Diagnosis
Will not log on
Verify that proper com port is selected on computer. (If a Serial to
USB adapter is used, find out the proper com port selection in the
“Device Manager”)
If error “Com not ready to send” check to see that only 1 session of
HVC Host is running. After correcting problem, it may be necessary
to reboot the computer.
Make sure that “Default Sandia=ID” in
“Host Setup” is set to 63.
Check Com cable
Check if pushbutton will scroll unit
Check battery voltage on LCD
Call factory for assistance.
Call factory for assistance
Will not scroll
Check battery voltage
Test scroll switch with ohmmeter
Restart instrument. Remove main power connector and then remove
the lithium jumper on the PCB. Wait 5 minutes, reconnect the main
power first (with good battery), then re-connect the lithium jumper
(No data will be lost after restart).
Call factory for assistance.
Display is blank
Check main & lithium battery voltage
Replace all batteries
Check lithium jumper (Scrape pins with a knife to clean if necessary)
Restart instrument. Remove and replace batteries and reconnect.
Call factory for assistance.
Not responding to
applied pressure
Confirm pressure mode set to AUTOCheck connections of transducer
wires
Check connections of transducer wires
Check “Full Scale” value in poly file. Calibrate column/ “Edit
Calibration Terms”, enter the correct “Full Scale”
Set the pressure calibration to default. -> Calibrate Column -> Edit
Calibration Terms -> click the check box next to Pressure Calibration
“Default”. Note: Recalibrate Pressure after these steps.
Replace transducer and recalibrate
Call factory for assistance