Typical Electrode Life

Ambient Temperature

1 – 3 years

90 ºC

Less than 4 months

120 ºC

Less than 1 month

Alkaline Error
High concentrations of sodium ions interfere with readings in alkaline

solutions. The pH at which the interference starts to be significant

depends upon the composition of the glass. This interference is called

alkaline error and causes the pH to be underestimated. Hanna’s glass

formulations have the indicated characteristics.

The resistance of glass electrodes partially depends on the temperature.
The lower the temperature, the higher the resistance. It takes more time
for the reading to stabilize if the resistance is higher. In addition, the
response time will suffer to a greater degree at temperatures below 25 ºC.

Since the resistance of the pH electrode is in the range of 50 – 200

Mohm, the current across the membrane is in the pico Ampere range.

Large currents can disturb the calibration of the electrode for many

hours.
For these reasons high humidity environments, short circuits and static

discharges are detrimental to a stable pH reading.
The pH electrode’s life also depends on the temperature. If constantly

used at high temperatures, the electrode life is drastically reduced.

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

TEMPERATURE CORRELATION

FOR

FOR

FOR

FOR

FOR pppppH SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

H SENSITIVE GLASS

1.0 Mol L

-1

Na

+

0.1 Mol L

-1

Na

+

Sodium Ion Correction for the Glass at 20-25 ºC

Concentration

pH

Error

13.00

13.50

14.00

12.50

13.00

13.50

14.00

0.10

0.14

0.20

0.10

0.18

0.29