Hanna Instruments HI 2221 User Manual
Page 36

36
CLEANING PROCEDURE
CLEANING PROCEDURE
CLEANING PROCEDURE
CLEANING PROCEDURE
CLEANING PROCEDURE
• General
Soak in Hanna HI 7061 or HI 8061 General Cleaning
Solution for approximately ½ hour.
• Protein
Soak in Hanna HI 7073 or HI 8073 Protein Cleaning
Solution for 15 minutes.
• Inorganic
Soak in Hanna HI 7074 Inorganic Cleaning Solution
for 15 minutes.
• Oil/grease
Rinse with Hanna HI 7077 or HI 8077 Oil and Fat
Cleaning Solution.
IMPORTANT: After performing any of the cleaning procedures, rinse the
electrode thoroughly with distilled water, refill the reference chamber
with fresh electrolyte (not necessary for gel-filled electrodes) and soak the
electrode in HI 70300 or HI 80300 Storage Solution for at least 1 hour
before taking measurements.