The instrument will answer for these commands with:
> where: COMMANDS REQUIRING AN ANSWER where the checksum is the bytes sum of the answer string sent as 2 ASCII Causes the instrument to send a complete set of readings in • pH, temperature and mV reading on pH range. The answer string contains: • Meter mode (2 chars): • 00 - pH value (0.001 resolution) • Meter status (2 chars of status byte): represents a 8 bit hexadecimal encoding. • Reading status (2 chars): R - in range, O - over range, U - under range. First character corresponds • pH reading on pH range only - 7 ASCII chars, including sign and decimal point. (Not for HANNA • mV reading - 7 ASCII chars, including sign and decimal point.
The instrument will answer for these commands with:
characters.
All the answer messages are with ASCII characters.
RAS
according with the current range:
• mV and temperature reading on mV range
• 01 - pH range (0.01 resolution)
• 03 - mV range
• 0x10 - temperature probe is connected
• 0x01 - new GLP data available
• 0x02 - new SETUP parameter
to the pH range reading. (Not for HANNA ORP
electrodes with pin.) Second character corresponds
to mV reading.
ORP electrodes with pin.)