• Remove the cuvette.

• Add the content of one packet of

HI 93716-0

Bromine reagent. Replace

the cap and shake gently for about 20
seconds to dissolve most of the reagent.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 2 minutes and 30
seconds then just press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
bromine on the LCD.

INTERFERENCES
Chlorine
Iodine
Ozone
Oxidized forms of chromium and manganese