beautypg.com

Calibration, Operational guide – Hanna Instruments HI 8033 User Manual

Page 4

background image

7

6

The instrument should be re-calibrated at least once a month, or
after probe or battery replacement.
For more accurate results, it is recommended to use a calibration
solution with a conductivity value close to the range to be measured.
See the "Accessories" section for a wide selection of conductivity
solutions.

PROCEDURE FOR HI 8033

• Pour the conductivity calibration solution

(e.g. HI 7030) into a beaker while mak-
ing sure that the holes on the probe are
completely submerged. If possible, use
plastic beakers to minimize any EMC in-
terference.

• Immerse the conductivity probe and the

ChecktempC in the solution.

• Wait a couple of minutes for the thermal

equilibrium to be reached.

• Tap the probe on the bottom, then shake it while rotating to

make sure no air bubbles remain trapped in the sleeve.

• Measure the temperature of the calibration solution with the

ChecktempC, and adjust the °C knob accordingly.

• Turn the rotary knob to select the 19990 µS/cm range.
• Turn the k% calibration knob until the display shows the

conductivity reading at 25°C (see "Conductivity vs. temperature
chart "), e.g. 12880 µS/cm (= 12.88 mS/cm), and all
subsequent measurements will be compensated to 25°C (77°F).

• If you prefer to standardize the temperature compensation to

20°C (68°F), adjust the trimmer to read "11.67 mS" (see
"Conductivity vs. temperature chart "). All subsequent measure-
ments will be compensated to 20°C.

• The calibration is now complete and the meter is ready for use.

PROCEDURE FOR HI 933000

• Pour the conductivity calibration solution (e.g. HI 7030) into a

beaker while making sure that the holes on the probe are
completely submerged. If possible, use plastic beakers to minimize
any EMC interference.

HI 7030

°C

CALIBRATION

• Each meter is supplied complete with a 9V battery. Slide off the

battery compartment cover on the back of the meter and install
the battery while paying attention to its polarity (see "Battery
Replacement" section for details).

• Connect the probe to the meter by aligning the pins with the

socket and pushing the plug in. With HI 933000, tighten the
threaded ring.

• Make sure that the meter has been calibrated before taking any

measurements (see "Calibration" section for details).

• Immerse the conductivity probe into the

sample, making sure that the holes on
the shaft are completely submerged.
If possible, use plastic beakers to minimize
any EMC interference.

• Tap the probe lightly on the bottom of the beaker to remove any

air bubbles which may be trapped inside the sleeve.

• Turn the instrument on by pressing the ON/OFF key or by setting

the rotary switch (HI 8033) to the desired measurement range.

• For HI 8033:

Measure the solution temperature with a
ChecktempC or another accurate ther-
mometer, and set the temperature knob
to the measured value (e.g. 20°C).

• Select the appropriate measurement range.

Note: If the display shows only a "1", the

meter is out of range. Select the next
(higher) range.

• For HI 933000:

Before taking any measurement, wait a couple of minutes for the
temperature sensor to reach thermal equilibrium with the sample.
When the temperature of the sample is lower than 20°C or higher
than 30°C, allow more time for the thermal equilibrium to be
achieved.

• After measurements, switch the instrument off and clean the

probe (see "Probe Maintenance" section for details).

OPERATIONAL GUIDE

°C

This manual is related to the following products: