beautypg.com

Functional description & specifications of hi 8033 – Hanna Instruments HI 8033 User Manual

Page 3

background image

5

4

1) Liquid Crystal Display
2) ON/OFF key
3) Measurement range selection keys
4) Probe connector

Range

0.0 to 199.9 / 0 to 1999 µS/cm

0.00 to 19.99 / 0.0 to 199.9 mS/cm

Resolution

0.1 / 1 µS/cm ; 0.01 / 0.1 mS/cm

Accuracy

(@ 20°C / 68°F)

±1% FS (excluding probe error)

Typical EMC Deviation

±2% FS

Calibration

Manual, 1 point through trimmer

Temperature

Automatic, 10 to 40°C (50 to 104°F)

Compensation

with ß = 2%/°C

Probe (included)

HI 76302W with 1 m (3.3') cable

Environment

0 to 50°C (32 to 122°F); RH max 95%

Battery Type

1 x 9 V alkaline

Battery Life

Approx. 100 hours of continuous use

Dimensions

143 x 80 x 38 mm (5.6 x 3.2 x1.5")

Weight

360 g (13 oz.)

FUNCTIONAL DESCRIPTION &

SPECIFICATIONS OF HI 933000

1) Liquid Crystal Display
2) Rotary switch
3) Knob for manual temperature compensation
4) Calibration knob

mS

Conductivity meter

HI 8033

ins rum n s

°C

k%

µ

S

0-199,9

µ

S

0-1999

µ

S

0-19,990

salinity
0-19,990

ppm
CaCO

3

OFF

20

30

10

40

0

50

5

5

0

10

10

15

15

+

-

1

2

3

4

FUNCTIONAL DESCRIPTION &

SPECIFICATIONS OF HI 8033

Range

0.0 to 199.9 ; 0 to 1999 µS/cm

0.00 to 19.99 mS/cm

0 to 19990 ppm

Resolution

0.1 ; 1 µS/cm / 0.01 mS/cm / 10 ppm

Accuracy

(@20°C/68°F)

±1% FS (excluding probe error)

Typical EMC Deviation

±2 % FS

Calibration

Manual, 1 point

Temperature

Manual, 0 to 50°C (32 to 122°F)

Compensation

with ß = 2%/°C

Probe (included)

HI 76301W with 1 m (3.3') cable

Environment

0 to 50°C (32 to 122°F); RH max 95%

Battery Type

1 x 9 V alkaline

Battery Life

Approx. 100 hours of continuous use

Dimensions

185 x 82 x 47 mm (7.3 x 3.2 x1.9")

Weight

270 g (9.5 oz.)

This manual is related to the following products: