Hanna Instruments HI 9141 User Manual
Page 8

15
14
SETTING
DATE, TIME & PRINTING INTERVAL
Turn the instrument on by
pressing the ON/OFF key and
wait for the "Cond" message
to disappear.
Press the FUNCTION key twice and the dis-
play will show "F1".
Press the CAL key and the display will show
the date setting.
Press the CAL key again and the year will
start blinking on the secondary LCD.
Use the UP or DOWN keys to set the year.
When the correct year is selected, press the
CAL key. The month will start blinking.
CAL
DATE
DATE
UP
DOWN
ON/OFF
DATE
FUNCTION
taking into account the lower degree of oxy-
gen solubility in such conditions (see "Alti-
tude Compensation" and "Salinity Compen-
sation" sections for details).
Remember to set altitude and/or salinity be-
fore taking ppm measurements, and the meter
will automatically compensate for these fac-
tors.
TEMPERATURE READINGS
The supplied probe is provided with a built-in
temperature sensor.
Press the FUNCTION key to display the
measured temperature.
Allow the probe to reach the thermal equilib-
rium before taking any measurement. This
can take several minutes. The greater the
difference between the temperature at which
the probe was stored and the temperature of
the sample, the longer the time will be.
If the display shows "°C ---.-", the D.O. probe
is not properly connected. This can also indi-
cate a broken probe cable.
FUNCTION
°C
°C
CAL
CAL