PROBE INITIAL PREPARATION

Probe Preparation
All D.O. probes from Hanna Instrument are shipped dry. To
hydrate the probe and prepare it for use, connect it to the
meter and proceed as follows.

1. Remove the red & black plastic cap. This

cap is used for shipping purposes only and
can be thrown away.

2. Wet the sensor by soaking the bottom

(2.5 cm/1") of the probe in

HI 7041S

electrolyte solution for 5 minutes.

3. Rinse the membrane (

HI 76407A supplied with the

meter) with some electrolyte
while shaking it gently. Refill
with clean electrolyte.

4. Gently tap the sides of the

membrane with your finger
to ensure that no air bubbles
remain trapped inside.

To avoid damaging the
membrane, do not tap the
membrane directly on the
bottom.

5. Place the rubber O-Ring properly inside the membrane

cap.

6. With the sensor facing down, screw the cap clockwise.

Some electrolyte will overflow.

When not in use and during polarization, protect
the membrane with the supplied cap.

Shipping

cap

black

red

THEN SCREW

BACK ON

THEN TAP

FILL FIRST

SPECIFICATIONS

Range

0.0 to 19.9 mg/L (ppm)

-5.0 to 50.0°C (23.0 to 122.0°F)

Resolution

0.1 mg/L

0.1°C (1°F)

Accuracy

±1.5% FS

±0.2°C (±1°F)

excluding probe error

Calibration

Manual in 100% DO

Automatic in 0% DO

Temperature

Automatic, 0 to 50°C

Compensation

(32 to 122°F)

Probe

HI 76407/4, polarographic,

with 4 m (13') cable (included)

Battery Type

3 x 1.5V AAA

Life

approx. 1000 hours of use

Auto-off

After approx. 8 minutes

Environment

0 to 50°C (32 to 122°F);

RH max 100%

Dimensions

185 x 72 x 36 mm (7.3 x 2.8 x 1.4")

Weight

300 g (10.6 oz.)