Hanna Instruments HI 83213-2008 User Manual
Page 39
39
Copper LR
COPPER LOW RANGE
SPECIFICATIONS
Range
0 to 1000 μg/L
Resolution
1 μg/L
Accuracy
±10 μg/L ±5% of reading at 25 °C
Typical EMC
±1 μg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 575 nm
Method
Adaptation of the
EPA method. The reaction between copper and the bicinchoninate
reagent causes a purple tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 95747-0
Bicinchoninate
1 packet
REAGENT SETS
HI 95747-01 Reagents for 100 tests
HI 95747-03 Reagents for 300 tests
For other accessories see page 72.
MEASUREMENT PROCEDURE
• Select the
Copper LR method using the procedure described in
the
Method Selection section (see page 12).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark) and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.
10 mL