Color of Water

COLOR OF WATER

SPECIFICATIONS

Range

0 to 500 PCU (Platinum Cobalt Units)

Resolution

1 PCU

Accuracy

±10 PCU ±5% of reading at 25 °C

Typical EMC

± 1 PCU

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 420 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, Colorimetric Platinum Cobalt method.

REQUIRED ACCESSORIES

0.45 μm membrane for true color measurement.
For other accessories see page 72.

MEASUREMENT PROCEDURE

• Select the

Color of Water method using the procedure described in

the

Method Selection section (see page 12).

• Fill one cuvette up to the mark with deionized water and replace

the cap. This is the blank.

• Place the blank (# 1) into the holder and close the lid.

• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the blank.

# 1

# 1

10 mL