Hanna Instruments HI 83214 User Manual
Page 42

42
• Select
Oxygen Demand, (COD) LR method following one of the procedures described in the “Method
Selection” section (see page 15).
• Place the blank vial into the holder and push it completely down.
• Press the Zero key. The instrument will perform a zero sequence and after a few seconds the display
will show “
-0.0-”. Now the meter is zeroed and ready for measurement.
• Remove the blank vial.
• Place the sample vial into the holder and push it completely down.
• Press
Read. The instrument will perform the reading.
• The instrument displays concentration in
mg/L of oxygen demand.
INTERFERENCES
Interference may be caused by:
Chloride (Cl-) above 2000 mg/L.
Samples with higher chloride concentration should be diluted.
COD LR
See also other documents in the category Hanna Instruments Measuring instruments:
- pH 213 (17 pages)
- pH 211R (13 pages)
- pH 21 (9 pages)
- pH 209V (9 pages)
- pH 209R (13 pages)
- HI 99181 (2 pages)
- HI 99171 (2 pages)
- HI 99163N (2 pages)
- HI 99161N (2 pages)
- HI 99141 (2 pages)
- HI 99131 (2 pages)
- HI 99121 (9 pages)
- HI 99104 (2 pages)
- HI 991003 (2 pages)
- HI 991002 (2 pages)
- HI 991001 (2 pages)
- HI 98402 (9 pages)
- HI 98401 (8 pages)
- HI 98308 (2 pages)
- HI 98240 (21 pages)
- HI 98203 (2 pages)
- HI 98202 (2 pages)
- HI 98201 (2 pages)
- HI 98185 (72 pages)
- HI 98172 (52 pages)
- HI 98160N (48 pages)
- HI 98160 (48 pages)
- HI 98150 (17 pages)
- HI 9815 (2 pages)
- HI 981410 (2 pages)
- HI 981408 (2 pages)
- HI 98121 (2 pages)
- HI 98120 (2 pages)
- HI 98113 (2 pages)
- HI 98110 (2 pages)
- HI 98108 (2 pages)
- HI 98107 (2 pages)
- HI 98106 (2 pages)
- HI 98103 (2 pages)
- HI 96108 (2 pages)
- HI 96107 (2 pages)
- HI 96106 (2 pages)
- HI 96104 (28 pages)
- HI 96101 (36 pages)
- HI 9321 (36 pages)