Hanna Instruments HI 83214 User Manual
Page 12

12
DIGESTION
• Some analytical methods require digestion of the sample. Use the
HI 839800 test tube heater for
digestion. The optional
HI 740217 safety shield is strongly recommended. For correct use of the reactor
follow the Reactor Instruction Manual. At the end of the digestion period the vials are still hot, allow
the vials to cool to room temperature in the optional
HI 740216 test tube cooling rack.
REAGENT BLANK CORRECTION
• Some methods require a “reagent blank correction”. The blank and the sample are prepared exactly
in the same way, with the only difference that for the blank deionized water is used instead of sample.
A blank vial may be used more than once; stability and storing conditions are described for each
method in the related chapter.
INTERFERENCES
• In the method related measurement sections we have reported the most common interferences that may
be present in an average wastewater matrix. It may be that for a particular treatment process other
compounds do interfere with the method of analysis.
HEALTH & SAFETY
• The chemicals contained in the reagent kits may be hazardous if improperly handled.
• Read the appropriate Material Safety Data Sheet (MSDS) before performing tests.
• Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions
carefully.
• Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water.
If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors.
• Reagent vial disposal: Reagents vials may contain different waste pollutants. After use dispose of the
reagent vials according to the local regulations.