Hanna Instruments HI 83218-01 User Manual
Page 30
![background image](/manuals/737577/30/background.png)
30
Phosphorus
PHOSPHORUS
SPECIFICATIONS
Range
0.0 to 15.0 mg/L
Resolution
0.1 mg/L
Accuracy
±0.3 mg/L ±4% of reading at 25 °C
Typical EMC Dev. ±0.2 mg/L
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the
Standard Methods for the Examination of Water and Wastewater,
18
th
edition, Amino Acid method. The reaction between phosphate and reagents
causes a blue tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93706A-0
Molybdate
10 drops
HI 93706B-0
Amino Acid Powder
1 packet
REAGENT SETS
HI 93706-01 Reagents for 100 tests
HI 93706-03 Reagents for 300 tests
For other accessories see page 34.
MEASUREMENT PROCEDURE
• Select the
Phosphorus method using the procedure described in
the
Method Selection section (see page 11).
• Fill the cuvette with 10 mL of unreacted sample (up to the mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.
• Remove the cuvette.
• Add 10 drops of HI 937O6A-0 Molybdate reagent.
10 mL