Hanna Instruments HI 83226-01 User Manual
Page 8
14
15
METHOD REFERENCE TABLE
Method
Method
Page
description
1
Alkalinity
20
2
Bromine
22
3
Calcium Hardness
24
4
Free Chlorine
27
5
Total Chlorine
29
6
Free Copper
31
HEALTH & SAFETY
• The chemicals contained in the reagent kits may be hazardous if improperly handled.
• Read the Material Safety Data Sheet (MSDS) before performing tests.
• Safety equipment: Wear suitable eye protection and clothing when required, and follow instructions
carefully.
• Reagent spills: If a reagent spill occurs, wipe up immediately and rinse with plenty of water.
If reagent contacts skin, rinse the affected area thoroughly with water. Avoid breathing released vapors.
• Waste disposal: for proper disposal of reagent kits and reacted samples, refer to the Material Safety
Data Sheet (MSDS).
Method
Method
Page
description
7
Total Copper
33
8
Cyanuric Acid
35
9
Iron
37
10
Ozone
39
11
pH
42
specific method.
• It is possible to take multiple readings in a row, but it is recommended to take a new zero reading for
each sample and to use the same cuvette for zeroing and measurement when possible (for most precise
results follow the measurement procedures carefully).
• Discard the sample immediately after the reading is taken, or the glass might become permanently
stained.
• All the reaction times reported in this manual are at 25 °C (77 °F). In general, the reaction time
should be increased for temperatures lower than 20 °C (68 °F), and decreased for temperatures higher
than 25 °C (77 °F).
INTERFERENCES
• In the method measurement section the most common interferences that may be present in an average
sample matrix have been reported. It may be that for a particular treatment process other compounds
do interfere with the method of analysis.
POWER CONNECTION AND BATTERY MANAGEMENT
The meter can be powered from an AC/DC adapter (included) or from the built-in rechargeable battery.
Note: Always turn the meter off before unplugging it to ensure no data is lost.
When the meter switches ON, it verifies if the power supply adapter is connected. The battery icon on the
LCD will indicate the battery status:
- battery is charging from external adapter
- battery fully charged (meter connected to AC/DC adapter)
- battery capacity (no external adapter)
- battery Low (no external adapter)
- battery Dead (no external adapter)
OPERATIONAL GUIDE
• Turn the instrument ON via the ON/OFF power switch.
• The meter will perform an autodiagnostic test. During this test, the Hanna Instrument logo will appear
on the LCD. After 5 seconds, if the test was successful, the last method used will appear on the display.
• In order to select the desired method press the METHOD key and a screen with the available methods
will appear.
• Press the ▲ ▼ keys to highlight the desired method. Press Select.
METHOD SELECTION