SPECIFICATIONS
Range
6.5 to 8.5 pH
Resolution
0.1 pH
Accuracy
±0.1 pH at 25 °C
Typical EMC
±0.1 pH
Deviation
Light Source
Tungsten lamp with narrow band interference filter @ 525 nm
Method
Adaptation of the Phenol Red method. The reaction with the reagent causes a yellow
to red tint in the sample.
REQUIRED REAGENTS
Code
Description
Quantity
HI 93710-0
Phenol Red Indicator
5 drops
REAGENT SETS
HI 93710-01 Reagents for 100 pH tests
HI 93710-03 Reagents for 300 pH tests
For other accessories see page 46.
MEASUREMENT PROCEDURE
• Select the
pH method using the procedure described in the
Method
Selection section (see page 15).
• Fill the cuvette with 10 mL of unreacted sample (up to the
mark)
and replace the cap.
• Place the cuvette into the holder and close the lid.
• Press the
Zero key. The display will show “-0.0-” when the meter is zeroed and ready for
measurement.
pH
• Remove the cuvette and add 5 drops of HI 93710-0 Phenol
Red Indicator. Replace the cap and mix the solution.
• Reinsert the cuvette into the instrument.
• Press the
Read key to start the reading. The instrument displays the pH value.
10 mL