beautypg.com

Hanna Instruments HI 95711 User Manual

Page 5

background image

5

Range

Free Cl

2

0.00 to 5.00 mg/L

Total Cl

2

0.00 to 5.00 mg/L

Resolution

0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L

Precision

±0.02 mg/L @ 1.00 mg/L

Typical EMC Deviation ±0.01 mg/L
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Light Detector

Silicon Photocell

Method

Adaptation of the USEPA method 330.5 and Standard Method 4500-Cl G.
The reaction between chlorine and the DPD reagent causes a pink tint in
the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode.

Dimensions

180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Weight

290 g (10 oz.).

REQUIRED REAGENTS

Code

Unit

Description

Quantity/test

HI 93701-0

Free Cl

2

Free Chlorine Reagent

1 packet

HI 93711-0

Total Cl

2

Total Chlorine Reagent

1 packet

Precision is how closely repeated
measurements agree with each other.
Precision is usually expressed as standard
deviation (SD).
Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good

accuracy, precise results can be inaccurate.
The figure explains these definitions. In a
laboratory using a standard solution of
1.00 mg/L chlorine and a representative lot of
reagent, an operator obtained with a single
instrument a standard deviation of 0.02 mg/L.

SPECIFICATIONS

PRECISION AND ACCURACY

20

REAGENT SETS
HI 93701-01

Reagents for 100 Free Chlorine tests

HI 93701-03

Reagents for 300 Free Chlorine tests

HI 93711-01

Reagents for 100 Total Chlorine tests

HI 93711-03

Reagents for 300 Total Chlorine tests

OTHER ACCESSORIES
HI 95701-11

CAL CHECK™ Standard Cuvets for Free Chlorine (1 set)

HI 95711-11

CAL CHECK™ Standard Cuvets for Total Chlorine (1 set)

HI 710009

Blue rubber boot

HI 710010

Orange rubber boot

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731331

Glass cuvets (4 pcs)

HI 731335

Caps for cuvets (4 pcs)

HI 93703-50 Cuvets cleaning solution (230 mL).

Battery replacement must only take place in a
non-hazardous area.
Simply slide off the battery cover on the back of
the meter. Detach the battery from the terminals
and attach a fresh 9V battery while paying attention
to the correct polarity. Insert the battery and replace
the cover.

9V

SLIDE

OFF

BATTERY REPLACEMENT