14

• Replace the cuvet into the holder and ensure

that the notch on the cap is positioned securely
into the groove.

• Hold READ/TIMER

for three seconds.

The

display will show the countdown prior to
measurement.
Alternatively, wait for 3 minutes and 30 seconds
and just press READ/TIMER.
In both cases “SIP” will blink during
measurement.

• The instrument directly displays concentration

in mg/L of ammonia nitrogen (NH

3

-N) on the

Liquid Crystal Display.

• To convert the reading to mg/L of ammonia

(NH

3

), multiply by a factor of 1.216.

or

INTERFERENCES

Organic compounds like: chloramines, various aliphatic and aromatic amines, glycine,
(positive error).
To eliminate these interferences distillation is required.

Organic compounds like: aldehydes, alcohols (e.g. ethanol) or acetone (negative error).
To eliminate these interferences distillation is required.

Sulfide (S

2-

): may cause turbidity.

Hardness above 1 g/L as Calcium Carbonate (CaCO

3

).