beautypg.com

Hanna Instruments HI 95715 User Manual

Page 12

background image

12

• It is important that the sample does not contain

any debris. This would corrupt the reading.

• In order to avoid reagent leaking and to obtain

more accurate measurements, it s recommended

to close the cuvet first with the supplied HDPE

plastic stopper and then with the black cap.

• Each time the cuvet is used, the cap must be

tightened to the same degree.

• Whenever the cuvet is placed into the measurement

cell, it must be dry outside, and completely free

of fingerprints, oil or dirt. Wipe it thoroughly with

HI 731318 (tissue for wiping cuvets, see chapter

ACCESSORIES) or a lint-free cloth prior to insertion.

• Do not let the reacted sample stand too long after

reagent is added, or accuracy will be lost.

• It is possible to take multiple readings in a row,

but it is recommended to take a new zero reading

for each sample and to use the same cuvet for

zeroing and measurement.

• After the reading it is important to discard

immediately the sample, otherwise the glass might
become permanently stained.

• All the reaction times reported in this manual are

referred to 20°C (68°F). As a general rule of

thumb, they should be doubled at 10°C (50°F)
and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a

measurement follow the validation procedure

to be sure that the instrument is properly

calibrated. If necessary, calibrate the instrument.

This manual is related to the following products: