Hanna Instruments HI 95729 User Manual

Page 5

background image

5

Range

0.00 to 2.00 mg/L

Resolution

0.01 mg/L

Precision

±0.03 mg/L @ 1.00 mg/L

Typical EMC Deviation

±0.01 mg/L

Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Light Detector

Silicon Photocell

Method

Adaptation of the

EPA method 340.1 and Standard Methods for

the Examination of Water and Wastewater, 20

th

edition, SPADNS

method. The reaction between fluoride and the liquid reagent
causes a red tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in

measurement mode;

after 1 hour of non-use in

calibration mode.

Dimensions

180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Weight

290 g (10 oz.).

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 93729-0

Fluoride Reagent

4 mL (2 mL x 2)

Precision is how closely repeated measurements
agree with each other. Precision is usually expressed
as standard deviation (SD).
Accuracy is defined as the nearness of a test result to
the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure explains
these definitions.
In a laboratory using a standard solution of 1.00
mg/L fluoride and a representative lot of reagents,
an operator obtained with a single instrument a
standard deviation of 0.03 mg/L.

SPECIFICATIONS

PRECISION AND ACCURACY

16

or

Note it is possible to interrupt calibration procedure at any time by pressing ON/OFF.
Warning: do not validate or calibrate the instrument with standard solutions other than the Hanna
CAL CHECK

TM

Standards, otherwise erroneous results will be obtained. For accurate Validation and

Calibration results, please perform test at room temperature: 18 to 25°C (64.5 to 77.0°F).

CALIBRATION PROCEDURE
• Turn the meter on by pressing ON/OFF.

When the LCD displays “729”, it is ready.

• Enter the

calibration mode by holding CAL CHECK

for three seconds

.

• The date of the last calibration appears (e.g.:

month “01”, day “08”). “F.CAL” means that
the factory calibration is selected.
Note: at this point it is possible to reset the
instrument to restore FACTORY CALIBRATION,
see instructions, page 18.

• Place the CAL CHECK

TM

Standard HI 95729-11

Cuvet

A into the holder and ensure that the

notch on the cap is positioned securely into
the groove.

• Press ZERO and “SIP” will blink on the display.

• After a few seconds the display will show

“-0.0-”. The meter is now zeroed and ready for
calibration.

• Remove the cuvet.

• Place the CAL CHECK

TM

Standard HI 95729-11

Cuvet

B into the holder and ensure that the

notch on the cap is positioned securely into
the groove.

CALIBRATION PROCEDURE