Hanna Instruments HI 95734 User Manual

Page 5

background image

5

Range

Free Cl

2

0.00 to 10.00 mg/L

Total Cl

2

0.00 to 10.00 mg/L

Resolution

0.01 mg/L from 0.00 to 3.50 mg/L; 0.10 mg/L above 3.50 mg/L

Precision

±0.06 mg/L @ 3.00 mg/L

Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Light Detector

Silicon Photocell

Method

Adaptation of the USEPA method 330.5 and Standard Method

4500-Cl G. The reaction between chlorine and the DPD reagent
causes a pink tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in measurement mode;
after 1 hour of non-use in calibration mode.

Dimensions

180 x 83 x 46 mm (7.1 x 3.3 x 1.8")

Weight

290 g (10 oz.).

REQUIRED REAGENTS

Code

Description

Quantity/test

HI 93701-0

Free Chlorine

Reagent

1 packet

HI 93734B-0

Free & Total Chlorine HR

Reagent

5 mL

HI 93734C-0

Total Chlorine HR Reagent

3 drops

Precision is how closely repeated
measurements agree with each other.
Precision is usually expressed as standard
deviation (SD). Accuracy is defined as the

nearness of a test result to the true value.
Although good precision suggests good
accuracy, precise results can be inaccurate. The
figure explains these definitions.
In a laboratory using a standard solution of
3.00 mg/L chlorine and a representative lot of
reagent, an operator obtained with a single
instrument a standard deviation of 0.06 mg/L.

PRECISION AND ACCURACY

SPECIFICATIONS

20

REAGENT SETS
HI 93734-01

Reagents for 100 Free or Total Chlorine HR tests

HI 93734-03

Reagents for 300 Free or Total Chlorine HR tests

OTHER ACCESSORIES
HI 95734-11

CAL CHECK™ Standard Cuvets (1 set)

HI 710009

Blue rubber boot

HI 710010

Orange rubber boot

HI 721310

9V battery (10 pcs)

HI 731318

Tissue for wiping cuvets (4 pcs)

HI 731331

Glass cuvets (4 pcs)

HI 731335

Caps for cuvets (4 pcs)

HI 740157

Plastic refilling pipette (20 pcs)

HI 740226

5 mL graduated syringe

HI 93703-50

Cuvets cleaning solution (230 mL)

or

FACTORY CALIBRATION RESET
It is possible to restore factory calibration:
• Enter the calibration mode by holding CAL CHECK

for three seconds.

• Hold ZERO/SET for 10 seconds. The display will

show for 2 seconds “F.CAL” and the parameter
code ("734f" or “734t”) appears. The factory
calibration is automatically restored and the
instrument is ready for measurement.

ACCESSORIES