• Press CAL CHECK and the lamp, cuvette and

detector icons together with “CAL CHECK”
will appear on the display, depending on the
measurement phase.

• At the end of the measurement the display

will show the validation standard value.

The reading should be within specifications as reported in the CAL CHECK

TM

Standard

Certificate. If the value is found out of the specifications, please check that the cuvettes are free
of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications,
then recalibrate the instrument.