Hanna Instruments HI 96704 User Manual

Page 13

background image

13

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.
The blinking “ZERO” indicates that the
instrument needs to be zeroed first.

• Fill one cuvette with 10 mL of unreacted

sample, up to the mark.

• Fill a second cuvette with 10 mL of

distilled water, up to the mark (this is
the blank).

• Add 12 drops of the HI 93704-0 reagent

to each cuvette. Replace the caps and
swirl the solutions.

• Place the blank into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 12 minutes then
just press ZERO/CFM.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

10 mL
sample

10 mL
blank

12

3

6

9

12’