Hanna Instruments HI 96707C User Manual
Page 18
18
• After a few seconds the display will show
“-0.0-”
. The meter is now zeroed and
ready for calibration. The blinking
“READ”
asks for reading calibration
standard.
• Remove the cuvette.
• Place the CAL CHECK
TM
Standard
HI 96707-11
Cuvette B into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
• Press READ/ /TIMER and the lamp,
cuvette and detector icons will appear on
the display, depending on the
measurement phase.
• After measurement the instrument will
show for three seconds the CAL CHECK
TM
Standard value.
Note:
If the display shows “STD HIGH”, the
standard value was too high. If the
display shows “STD LOW”, the
standard value was too low. Verify
that both CAL CHECK
TM
Standards
HI 96707-11
Cuvettes, A and B are
free from fingerprints or dirt and that
they are inserted correctly.
• Then the date of the last calibration
(e.g.: ”01.08.2009”) appears on the
display, or “01.01.2009” if the factory
calibration was selected before. In both
cases the year number is blinking, ready
for date input.