Validation procedure – Hanna Instruments HI 96707C User Manual

Page 15

background image

15

INTERFERENCES

Highly buffered samples (alkalinity above 500 ppm) or with extreme pH may exceed the
buffering capacity of the reagent: neutralize acidic samples with diluted NaOH or alkaline
samples with diluted HCl, prior to addition of the reagent.
Turbidity must be removed by prior filtration.
The following ions may cause negative errors:
cupric ions (Cu

2+

):

above 5 ppm

ferrous ions (Fe

2+

): above 1 ppm

ferric ions (Fe

3+

):

above 2 ppm

VALIDATION PROCEDURE

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained. For accurate validation, please

perform test at room temperature: 18 to 25°C (64.5 to 77.0°F)

• Turn the meter on by pressing ON/OFF.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.

• Place the CAL CHECK

TM

Standard

HI 96707-11

Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.