Calibration procedure – Hanna Instruments HI 96715C User Manual
Page 9
![background image](/manuals/737487/9/background.png)
16
17
CALIBRATION PROCEDURE
Note:
It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or
ON/OFF
keys.
Warning
: Do not calibrate the instrument with standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained. For accurate calibration, please
perform test at room temperature: 18 to 25°C (64.5 to 77.0°F)
• After a few seconds, the display will show
“-0.0-”
. The meter is now zeroed and ready
for validation.
• Remove the cuvette.
• Place the CAL CHECK
TM
Standard HI 96715-11
Cuvette B into the cuvette holder and ensure
that the notch on the cap is positioned securely
into the groove.
• Press CAL CHECK
TM
and the lamp, cuvette
and detector icons together with “CAL CHECK”
will appear on the display, depending on the
measurement phase.
• At the end of the measurement the display
will show the validation standard value.
The reading should be within specifications as reported in the CAL CHECK
TM
Standard
Certificate. If the value is found out of the specifications, please check that the cuvettes are free
of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications,
then recalibrate the instrument.
• Turn the meter on by pressing ON/OFF.
• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
• Press and hold CAL CHECK
for three
seconds to
enter
calibration mode. The
display will show “CAL” during
calibration procedure. The blinking
“ZERO”
asks for instrument zeroing.
• Place the CAL CHECK
TM
Standard
HI 96715-11
Cuvette A into the cuvette
holder and ensure that the notch on the
cap is positioned securely into the groove.
• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
• Press ZERO/CFM and the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
• Place the CAL CHECK
TM
Standard HI 96715-11
Cuvette A into the cuvette holder and ensure that
the notch on the cap is positioned securely
into the groove.