Hanna Instruments HI 96715C User Manual
Page 8
![background image](/manuals/737487/8/background.png)
14
15
• Press ZERO/CFM and the lamp, cuvette
and detector icons will appear on the
display, depending on the measurement
phase.
• After a few seconds, the display will show
“-0.0-”
. The meter is now zeroed and
ready for measurement.
• Remove the cuvette, open it and add 4
drops of HI 93715A-0 First Reagent (6
drops in case of seawater analysis). Swirl
the cuvette gently to mix.
• Add 4 drops of HI 93715B-0 Second
Reagent to the cuvette (10 drops in case
of seawater analysis). Replace the cap
and swirl the cuvette to mix.
• Replace the cuvette into the cuvette holder
and ensure that the notch on the cap is
positioned securely into the groove.
• Press and hold READ/
READ/
READ/
READ/
READ/ /TIMER
/TIMER
/TIMER
/TIMER
/TIMER
for three
seconds.
The display will show the
countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 3 minutes and 30
seconds and just press READ/
READ/
READ/
READ/
READ/ /TIMER
/TIMER
/TIMER
/TIMER
/TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.
• At the end of measurement, the instrument
directly displays concentration in mg/L of
ammonia nitrogen (NH
3
-N) on the LCD.
• To convert the reading to mg/L of
ammonia (NH
3
), multiply by a factor of
1.216.
VALIDATION PROCEDURE
Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:
Do not validate the instrument with any standard solutions other than the HANNA CAL
CHECK
TM
Standards, otherwise erroneous results will be obtained. For accurate validation, please
perform test at room temperature: 18 to 25°C (64.5 to 77.0°F)
•
Turn the meter on by pressing ON/OFF.
• When the beeper sounds briefly and the
LCD displays dashes, the meter is ready.
INTERFERENCES
•
Organic compounds like: chloramines, various aliphatic and aromatic amines, glycine,
(positive error).
To eliminate these interferences distillation is required.
•
Organic compounds like: aldehydes, alcohols (e.g. ethanol) or acetone (negative error).
To eliminate these interferences distillation is required.
•
Sulfide (S
2-
): may cause turbidity.
•
Hardness above 1 g/L as Calcium Carbonate (CaCO
3
).