5

SPECIFICATIONS

PRECISION AND ACCURACY

Range

0.0 to 12.5 mg/L

Resolution

0.1 mg/L

Accuracy

±0.1 mg/L ±5% of reading @ 25°C

Typical EMC Deviation

±0.1 mg/L

Light Source

Tungsten lamp

Light Detector

Silicon Photocell with narrow band interference filter @ 525nm

Method

Adaptation of the Standard Methods for the Examination of Water and
Wastewater, 18

th

edition, DPD method. The reaction between iodine

and the reagent causes a pink tint in the sample.

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Battery Type

1 x 9 volt

Auto-Shut off

After 10' of non-use in

measurement mode

;

after 1 hour of non-use in

calibration mode

;

with last reading reminder.

Dimensions

192 x 102 x 67 mm (7.6 x 4 x 2.6")

Weight

290 g (10 oz.).

REQUIRED REAGENTS
Code

Description

Quantity/test

HI 93718-0

DPD Powder Reagent

1 packet

Precision

Precision

Precision

Precision

Precision is how closely repeated measurements
agree with each other. Precision is usually
expressed as standard deviation (SD).
Accuracy

Accuracy

Accuracy

Accuracy

Accuracy is defined as the nearness of a test
result to the true value.
Although good precision suggests good accuracy,
precise results can be inaccurate. The figure
explains these definitions.
In a laboratory using a standard solution of 2.5
mg/L iodine and a representative lot of reagent,
an operator obtained with a single instrument a
standard deviation of 0.1 mg/L.