VALIDATION PROCEDURE

Alkalinity above 250 mg/L and acidity above 150 mg/L as CaCO

3

require neutralization since the

color could instantly fade. To resolve this, neutralize the sample with diluted HCl or NaOH.
In case of water with hardness greater than 500 mg/L CaCO

3

, shake the sample for approxi-

mately 2 minutes after adding the reagent.

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained.

Turn the meter on by pressing ON/OFF.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• Place the CAL CHECK

TM

Standard

HI 96718-11

Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.