Validation procedure – Hanna Instruments HI 96751 User Manual

Page 8

background image

14

15

VALIDATION PROCEDURE

• Remove the cuvette.

• Add the content of one packet of

HI 93751-0

reagent, replace the cap

and shake gently the cuvette for 1 minute.

• Replace the cuvette into the cuvette holder

and ensure that the notch on the cap is
positioned securely into the groove.

• Press and hold READ/ /TIMER

for three

seconds.

The display will show the

countdown prior to measurement. The
beeper is playing a beep at the end of
countdown period.
Alternatively, wait for 5 minutes then just
press READ/ /TIMER.
In both cases, the lamp, cuvette and
detector icons will appear on the display,
depending on the measurement phase.

• At the end of measurement, the instrument

directly displays concentration in mg/L of
sulfate on the LCD.

Calcium (as CaCO

3

)

>20000 mg/L

Chloride (as Cl

-

)

>40000 mg/L

Magnesium (as MgCO

3

)

>10000 mg/L

Silica (as SiO

2

)

>500 mg/L

Use the validation procedure to ensure that the instrument is properly calibrated.
Warning:

Do not validate the instrument with any standard solutions other than the HANNA CAL

CHECK

TM

Standards, otherwise erroneous results will be obtained.

• Turn the meter on by pressing ON/OFF.

• Press ZERO/CFM and the lamp, cuvette

and detector icons will appear on the
display, depending on the measurement
phase.

• Place the CAL CHECK

TM

Standard

HI 96751-11

Cuvette A into the cuvette

holder and ensure that the notch on the
cap is positioned securely into the groove.

INTERFERENCES
Color or suspended matter in large amounts will interfere: suspended matter should be removed
by previous filtration.
Organic matter in large amounts may impede the precipitation of barium sulfate.

• When the beeper sounds briefly and the

LCD displays dashes, the meter is ready.
The blinking “ZERO” asks for instrument
zeroing.