beautypg.com

Measurement procedure startup – Hanna Instruments HI 96753 User Manual

Page 7

background image

12

13

• It is important that the sample does not contain any debris. This would corrupt the reading.
• Each time the cuvette is used, the cap must be tightened to the same degree.
• Whenever the cuvette is placed into the measurement cell, it must be

dry outside, and completely free of fingerprints, oil or dirt. Wipe it
thoroughly with HI 731318 or a lint-free cloth prior to insertion.

• Shaking the cuvette can generate bubbles in the sample, causing

higher readings. To obtain accurate measurements, remove such bubbles
by swirling or by gently tapping the cuvette.

• Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.
• It is possible to take multiple readings in a row, but it is recommended to take a new zero

reading for each sample and to use the same cuvette for zeroing and measurement.

• After the reading it is important to discard immediately the sample, otherwise the glass

might become permanently stained.

• All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule

of thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).

• In order to maximize accuracy, prior to a measurement follow the validation procedure to

be sure that the instrument is properly calibrated. If necessary, calibrate the instrument.

MEASUREMENT PROCEDURE

STARTUP

Prepare the instrument for measurement as follows:
• Unpack the instrument by removing the dust protection sleeve from the instrument cuvette

holder.

• Place the battery in the instrument as described in the “BATTERY REPLACEMENT” chapter.
• Place the instrument on a flat table.
• Do not place the instrument under direct sun light.

To compensate the meter for the sample turbidity or color, the measurement takes place in two
phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the
reacted sample is measured.

• When the beeper sounds briefly and the LCD

displays dashes, the meter is ready. The blinking
“ZERO”

indicates that the instrument needs to

be zeroed first.

• Fill one cuvette (#1) with 10 mL of distilled

water (up to the mark).

• Fill another cuvette (#2) with 10 mL of

sample (up to the mark)
Note: for low chloride ions concentration, rinse
the cuvette a couple of times with sample
before filling it with 10 mL of sample (up to the
mark).
Note: for most accurate results, use two gradu-
ated pipettes to deliver exactly 10 mL of distilled water and 10 mL of sample to the cuvettes.

• Using the 1 mL syringe, add 0.5 mL of HI 93753A-0 Displacing Reagent to each cuvette.
• Replace the caps and mix each cuvette by turning the cuvette upside down for approximately

30 seconds.

• Turn the meter on by pressing ON/OFF.

The display briefly shows all tags on.