beautypg.com

Hanna Instruments HI 96804 User Manual

Page 2

background image

2

3

All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the

copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA.

Remove the instrument from the packing materials and examine carefully to ensure no
damage has occurred during shipping. If any damage has occurred, notify your Dealer
or closest Hanna Customer Service Center.
Each HI 96804 instrument is supplied with:

• 9 V battery
• Instruction manual

Note:Save all packing material until you are sure that the instrument functions

correctly. A defective instrument must be returned in its original packing.

Dear Customer,
Thank you for choosing a Hanna product. This manual will provide you with the necessary

information for the correct use of the instrument. Please read it carefully before using the meter. If

you need additional technical information, do not hesitate to e-mail us at [email protected].
This instrument is in compliance with directives.

TABLE OF CONTENTS

PRELIMINARY EXAMINATION

PRELIMINARY EXAMINATION ............................................................................................. 2
GENERAL DESCRIPTION ..................................................................................................... 3
SPECIFICATIONS .................................................................................................................. 3
PRINCIPLE OF OPERATION ................................................................................................ 4
FUNCTIONAL DESCRIPTION................................................................................................ 5
DISPLAY ELEMENTS ............................................................................................................ 6
WARNINGS ........................................................................................................................... 6
ERROR MESSAGES .............................................................................................................. 7
CALIBRATION PROCEDURE ................................................................................................ 8
MEASUREMENT PROCEDURE ............................................................................................. 9
CHANGING TEMPERATURE UNITS ...................................................................................10
BATTERY REPLACEMENT .................................................................................................11
WARRANTY ........................................................................................................................11

MODEL IDENTIFICATION
HI 96804

GENERAL DESCRIPTION

The HI 96804 Digital Invert Sugar Refractometer is a rugged portable, water resistant device that
benefits from Hanna’s years of experience as a manufacturer of analytical instruments.
The HI 96804 is an optical instrument that employs the measurement of refractive index to
determine the % Invert Sugar in aqueous solutions. The method is both simple and quick. Samples
are measured after a simple user calibration with deionized or distilled water. Within seconds the
instrument measures the refractive index of the sample and converts it to % by weight concentration

units. The HI 96804 digital refractometer eliminates the uncertainity associated with mechanical
refractometers and is easily portable for measurements in the field.
The measurement technique and temperature compensation employ methodology recommended in
the ICUMSA Methods Book (Internationally recognized body for Sugar Analysis).
Temperature (in °C or °F) is displayed simultaneously with the measurement on the large dual level
display along with icons for Low Power and other helpful message codes.
Key features include:
• Dual-level LCD
• Automatic Temperature Compensation (ATC)
• Easy setup and storage
• Battery operation with Low Power indicator (BEPS)
• Automatically turns off after 3 minutes of non-use.

SPECIFICATIONS

Range:

0 to 85% / 0 to 80°C (32 to 176°F)

Resolution:

± 0.1% / ±0.1°C (±0.1°F)

Accuracy:

± 0.2% / ± 0.3°C (±0.5°F)

Temperature Compensation:

Automatic between 10 and 40°C (50 - 104°F)

Measurement Time:

Approximately 1.5 seconds

Minimum Sample Volume:

100μL (cover prism totally)

Light Source:

Yellow LED

Sample Cell:

SS ring and flint glass prism

Case Material:

ABS

Enclosure Rating:

IP 65

Battery Type / Life:

1 X 9V / 5000 readings

Auto-Off:

After 3 minutes of non-use

Dimensions:

19.2(W) x 10.2(D) x 6.7 (H)cm

Mass:

420g