- Air bubbles also disturb measurements and the probe should be

installed in such a way as to minimize them.

Note: It is always recommended to keep at least one spare probe

handy. When anomalies are not resolved with a simple mainte-
nance, change the electrode (and recalibrate the controller) to see
if the problem is alleviated.

Note: Ensure that the probe is installed in such a way that it perma-

nently lies in the solution whether in the tank or the circulation
pipe.

ACCESSORIES

pH electrodes
HI 1001

pH electrode for continuous flow-through monitoring

HI 1002

pH electrode for continuous flow-through monitoring

HI 1003

pH electrode with matching pin

EC probes
HI 7632/D

Conductivity probe, ATC, 1/2” thread, with DIN
connector and 2 m (6.6’) cable, for continuous flow-
through monitoring up to 3 atm pressure

HI 3003/D

Conductivity probe, ATC, Pt-ring, Kynar

®

body,

1/2” thread, with DIN connector and 2 m (6.6’)
cable, for continuous flow-through monitoring up to
6 atm pressure

pH calibration solutions
HI 7004L

pH 4.01 buffer solution, 500 mL bottle

HI 7007L

pH 7.01 buffer solution, 500 mL bottle

HI 7010L

pH 10.01 buffer solution, 500 mL bottle

EC calibration solutions
HI 7031L

1.41 mS/cm calibration solution, 500 mL bottle

HI 7039L

5.00 mS/cm calibration solution, 500 mL bottle

Maintenance solutions
HI 70300L

Electrode storage solution, 500 mL bottle

HI 7061L

General cleaning solution, 500 mL bottle

Other accessories
BL1.5, BL3, BL5, BL7, BL10, BL15, BL20

Dosing pumps with flow rate from 1.5 to 18.3 lph
(0.4 to 4.8 gph)

APPENDIX - A

pH VALUES AT VARIOUS TEMPERATURES.

The temperature directly affects the pH value.
For example, if the buffer’s temperature is 25°C (77°F), calibrate the
meter to read 7.01, 4.01 or 10.01; if the temperature is 20°C, calibrate
the meter to 7.03, 4.00 or 10.06; if the temperature is 50°C, calibrate
the meter to 6.98, 4.06 or 9.82; etc.
Please refer to the following chart for a more accurate pH calibration.

TEMP

pH VALUES

°C

°F

4.01

6.86

7.01

9.18

10.01

0
5

10
15
20
25
30
35
40
45
50
55
60
65
70

32
41
50
59
68
77
86
95

104
113
122
131
140
149
158

4.01
4.00
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
4.11
4.12

6.98
6.95
6.92
6.90
6.88
6.86
6.85
6.84
6.84
6.83
6.83
6.84
6.84
6.85
6.85

7.13
7.10
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
6.99
6.99

9.46
9.39
9.33
9.27
9.22
9.18
9.14
9.10
9.07
9.04
9.01
8.99
8.97
8.95
8.93

10.32
10.24
10.18
10.12
10.06
10.01

9.96
9.92
9.88
9.85
9.82
9.79
9.77
9.76
9.75