beautypg.com

Controller settings, Alarm conditions – Hanna Instruments HI 9914 User Manual

Page 12

background image

23

22

CONTROLLER SETTINGS

SETTING THE pH CONTROL

To adjust the pH setpoint:
• Press the SET button close to the pH display to read the setpoint value.

• Turn the SET knob until the display shows the desired value.
• Return to measurement mode by pressing the MEA button.

To select the dosing direction:
• Select the desired solution for pH correction by setting the ACID/BASE

switch.

• If the switch is in the ACID position, the pH pump is activated when

the pH value exceeds the setpoint. The controller will dose
acidic solution to reduce the pH until the user-selected
setpoint is reached.

• If the switch is in the BASE position, the pH pump is

activated when the pH value falls below the setpoint.
The controller will dose alkaline solution to increase the
pH until the user-selected setpoint is reached.

SETTING THE EC CONTROL
To adjust the Conductivity setpoint:
• Push the SET button close to EC display to read the setpoint value.

ALARM CONDITIONS

When an alarm condition is reached, the

ALARM LED turns ON and the

alarm relay contact is closed (short between the

COMMON and NO pins).

All the pumps and the water nozzle are off. The Conductivity and pH
measuring channels display the EC and pH values.
An alarm condition occurs when:
• the water remains below the low level (the sensor contact is open) for

about 30 seconds;

• the high (alarm) level is reached (the sensor contact is closed);
• the humidity sensor detects water leakage;
• an impossible condition is detected, for example when the low level is

not reached and the normal level is reached.

Reset button
To return to normal function, it is necessary that:
• the water level in the tank decreases

below the high (alarm) level sensor;

• the level sensors operate correctly;
• the humidity sensor is dry;
• the

RESET button is pressed.