Hanna Instruments HI 9931 User Manual

Page 8

background image

8

SETPOINT RANGE

From 0 to 10.00 mS/cm

Specifications

HI 9931

RANGE

0.00 to 10.00 mS/cm (EC)

RESOLUTION

0.01 mS/cm

ACCURACY

±2% F. S.

(@20 C/68 F)
TYPICAL EMC

±2% F. S.

DEVIATION

PROPORTIONAL

Conductivity from 0.0 to 1.6 mS/cm and

CONTROL

time cycle from 0 to 90 seconds

ALARM
CONTACTS

DOSING TERMINALS

PROBE (optional) 4-ring potentiometric with built-in temperature sensor
POWER SUPPLY 220/240V or110/115V at 50/60Hz

ENVIRONMENT

-10 to 50°C (14 to 122°F)

max. 95% RH non-condensing

WEIGHT

1.6 Kg (3.5 lb.)

ENCLOSURE

181 x 221 x 142mm (7.1 x 8.7 x 5.6")

CASE MATERIAL

Fiber-reinforced, self-extinguishing ABS

Terminal can be configured as normally open or
normally closed (isolated output Max. 2A, Max. 240V,
resistive load, 1,000,000 strokes). The alarm is
activated if conductivity exceeds by more than the user-
selectable interval (0 to 2.0 mS/cm) from the setpoint
or due to overdosage
Relay (115 to 240V, Max.2A,1,000,000 strokes) is
activated whenever conductivity falls below the setpoint

CALIBRATION Through "ZERO CAL" and "SLOPE CAL" trimmers