BATTERY REPLACEMENT

A “LOBAT” indication
appears on the display
when the batteries are
weak and require re-
placement. The instrument can still perform
approximately 50 measurements.

A “-BA-” indication will
appear when the bat-
teries are too weak to
ensure accurate mea-
surements. This mes-
sage appears for a few seconds and then the
meter turns off. Batteries must be replaced.

Batteries should only be replaced in a safe
area and using 1.5V AA alkaline type.

To replace the batter-
ies, simply remove the
two screws on the rear
cover of the instrument
and replace all four 1.5V
AA batteries with new
ones, while paying at-
tention to their polarity.

The meter will store new settings in its non-
volatile memory and the display will flash
“CAL” and “Stor” alternately for several sec-
onds.

USER-SELECTABLE SHUTDOWN

With HI 93114, the users can customize the
shutdown time to save power.

To change the shutdown
time, enter the diagnos-
tic mode by momentarily
pressing ALT and CAL.
Select mode 10 and press the

key repeat-

edly to set the desired shutdown time from
10 to 60 minutes with 10 minute increments,
or disable the shutdown mode by choosing
the OFF selection.

After the selection is
completed, exit the di-
agnostic mode by press-
ing ALT+CAL together.

To clear the buffer,
press ALT and CAL.

Select mode 40. Press
the

key. The display will show the “Cln”

message, indicating that memory is being
cleaned.

The lot number will be reset to 00 automatically.

REVIEWING THE CURRENT LOT NUMBER
To check the current
(vacant) lot number,
while in log mode,
press ALT and to-
gether.

+

+

+