Hanna Instruments HI 93124 User Manual

Page 4

background image

6

7

The instrument is supplied complete with batteries. Before starting operations, install
the batteries while paying attention to their polarity (see also “Battery Replacement”
section). To maximize the battery life the meter automatically switches off after 5
minutes of non-use. For reactivating it, simply press the ON/OFF key.

TAKING MEASUREMENTS

• Turn the meter ON by pressing the ON/OFF key.

• At start-up the meter will show the full display for

a few seconds, followed by the indication of the
remaining battery percentage.

• When the LCD displays “-- --” the meter is ready

to measure.

• Fill a clean cuvette up to 0.5 cm (one quarter inch)

from its rim with the thoroughly agitated sample.

• Allow sufficient time for bubbles to escape before securing the cap.

Note: Do not over-tighten the cap.

• Wipe the cuvette thoroughly with a lint-free tissue

(HI731318) before inserting into the measure-
ment cell.
The cuvette must be completely free of fingerprints
and other oil or dirt, particularly in the area where
the light goes through (approximately the bot-
tom 2 cm/1 inch of the cuvette).

• Place the cuvette into the cell and check that the

notch on the cap is positioned securely into the
groove.

• The mark on the cuvette cap should point towards

the LCD.

OPERATIONAL GUIDE

SPECIFICATIONS

Range

0 to 10.00 EBC

10 to 250 EBC

Resolution

0.01 EBC or 1 EBC

Accuracy ±0.20 EBC ±3% of reading (0.00 to 10.00 EBC)

±1 EBC ±5% of reading (outside)

Typical EMC Deviation

±1% F.S.

Light Source

Infrared LED @890 nm

Light Source Life

Life of the instrument

Light Detector

Silicon Photocell

Power Source

4 x 1.5V AA alkaline batteries

Battery Life

60 hours or 900 measurements

Auto-off

After 5 minutes of non-use

Environment

0 to 50

°

C (32 to 122

°

F);

RH max 95% non-condensing

Dimensions

220 x 82 x 66 mm (8.7 x 3.2 x 2.6")

Weight

510 g (1.1 lb.)