To prepare the instrument for taking
measurements, first install the batteries (see
Battery Replacement section on page 20) and
then turn the instrument on.

To maximize battery-life, the display is
automatically switched off after 5-minutes of
non-use. To reactivate the display, simply
press the ON/OFF key.

MEASUREMENT PROCEDURE:

Turn the meter on by

pressing the ON/OFF
key.

The meter will carry

out a self-test display-
ing a full set of figures.

The meter will carry

out a battery test show-
ing in percentage the
battery life left.

When the LCD dis-

plays “----” the meter
is ready to measure.

Fill a clean cuvet up to one quar-

ter inch (0.5 cm) from its rim with
the thoroughly agitated sample.

Allow sufficient time for bubbles

to escape before securing the cap.
Note: do not overtighten the cap.

LOGGING

MICROPROCESSOR

TURBIDITY METER

LOGGING

MICROPROCESSOR

TURBIDITY METER

HI 93703-11

VIEW

VIEW

STO

STO

CLR

CLR

GLP

GLP

CAL

CAL

DATE

DATE

READ

READ

ALT

ALT

OFF

ON

1/4“ 0.5

cm

%

ba

98

OPERATIONAL GUIDE

LOG

LOBAT

TIME

DATE %

188

17

If "ERR1" appearson

the LCD, please check
the standard solution.

After approximately 30 seconds the meter

will display 10.00, prompting the user to
place the 10.00 FTU standard solution in
the cuvet holder.

Place the 10.00 FTU

standard in the holder press CAL, SIP and
CL will start blinking

After approximately

30 seconds the meter
will display 500, asking the user to place
the 500 FTU buffer
solution in the cuvet
holder.

NOTE: At this point the

user can save the two point calibration
setup by pressing ALT & CAL buttons,
leaving the calibration mode.

To perform a three point
calibration, place the 500
FTU standard solution in the
cuvet holder.

Press CAL, SIP and

CL will start blinking

10.00

cl

sip

cl

500

cl

sip

cl

err1