Full, Err2 err3, Err5 – Hanna Instruments HI 93703 User Manual

Page 11: Err4, Real time clock error, Eeprom error, Internal communica- tion error, Internal bus error, Contact the hanna office nearest to you

background image

11

LOG-ON-DEMAND:

After taking a read-

ing, press the STO
button. The last
sample read will be
stored in memory.
The display will also
show on the bottom
right corner the reference number of the
stored sample.

NOTE: When the maximum number of samples

has been reached
(199), the LCD dis-
play will show "FULL".
This is a warning that
you will be overwrit-
ing the oldest sample recorded. To store the
current sample, press the STO button again
and this will overwrite the oldest sample in the
memory. All samples are shifted (reference
numbers reduced by one). Sample number
one remains the oldest sample.

CLEARING LOGGED DATA:

Press the ALT &

CLR buttons to-
gether. The display
will show "CLR" for
confirmation. Press
ALT & CLR again to
clear memory. To
abort the operation,
press any button.

ALL SAMPLES PREVIOUSLY STORED
WILL BE ERASED FROM MEMORY!

LOGGING

MICROPROCESSOR

TURBIDITY METER

LOGGING

MICROPROCESSOR

TURBIDITY METER

HI 93703-11

VIEW

VIEW

STO

STO

CLR

CLR

GLP

GLP

CAL

CAL

DATE

DATE

READ

READ

ALT

ALT

OFF

ON

FULL

LOGGING

MICROPROCESSOR

TURBIDITY METER

LOGGING

MICROPROCESSOR

TURBIDITY METER

HI 93703-11

VIEW

VIEW

STO

STO

CLR

CLR

GLP

GLP

CAL

CAL

DATE

DATE

READ

READ

ALT

ALT

OFF

ON

CLR

LOG

14

Real Time Clock

error

EEPROM error

Internal communica-

tion error

Internal Bus error

Contact the HANNA office nearest to you.

err2

err3

err4

err5

ENSURE ACCURATE MEASUREMENTS

Each time the cuvet is used, tighten the cap

to the same degree.

Discard the sample soon after the reading is

taken to avoid permanently clouding the glass.

All glassware used to contain the standards

and the samples should be maintained clean,
washed with the HI 93703-50 cleaning solution
and rinsed with HI 93703-0 or turbidity-free
water.

• Collect the samples in clean glass or plastic

bottles, fit stoppers and perform the analysis
quickly. If necessary, store the sample in a
cool, dark place, for up to a maximum of 24