Hanna Instruments HI 110 User Manual
Page 18

18
CALIBRATION ALARM TIME-OUT SET
Press CAL when the calibration time-out is displayed. Calibration time-out
(OFF or 1 to 7 days) will start blinking.
Press the ARROWkeys to change the calibration time-out value.
Press CFM to save the modified calibration time-out value.
Press CAL to escape without saving.
FIRST CUSTOM BUFFER SET
Press CAL when custom buffer 1 is displayed. The custom buffer (disabled no
cb1 or 0 to 16 pH) will start blinking.
Press the ARROW keys to change the custom buffer value.
Press CFM to save the modified custom buffer value.
Press CAL to escape without saving.
SECOND CUSTOM BUFFER SET
Press CAL when custom buffer 2 is displayed. The custom buffer (disabled no
cb2 or 0 to 16 pH) will start blinking.
Press the ARROW keys to change the custom buffer value.
Press CFM to save the modified custom buffer value.
Press CAL to escape without saving.
ONE-POINT CALIBRATION BEHAVIOUR
Press CAL when 1 Pnt message is displayed on the secondary LCD. One of
the two options (Pnt or OFFS) will start blinking (see pH CALIBRATION
PROCEDURE for details, page 11).
Press the ARROW keys to toggle between Pnt and OFFS options.
Press CFM to save the behaviour for one-point calibration.
Press CAL to escape without saving.
CAL
BUFFER pH
CAL
BUFFER pH
INTV
CAL
CAL