Setup – Hanna Instruments HI 110 User Manual
Page 17

17
Setup mode allows viewing and modifying the following parameters:
Calibration Alarm Time-Out
Custom buffer 1
Custom buffer 2
One-point calibration behaviour
Instrument ID (HI 112 & HI 113)
Current Time
Current Date
Beep Status
Baud Rate (serial communication, HI 112 & HI 113)
Command prefix (serial communication, HI 112 & HI 113)
To enter SETUP mode press SET while the instrument is in measurement mode.
Select a parameter with the ARROW keys.
Press CAL if you want to change a parameter value. The selected parameter
will start blinking.
Press RANGE (SEL for HI 110) to toggle between displayed parameters.
Press the ARROW keys to increase or decrease the displayed value.
Press CFM to save the modified value or CAL to escape.
Press the ARROW keys to select the next/previous parameter.
INTV
CAL
CAL
or the number of days until the calibration alarm will be displayed.
or if expired (7 days ago).
The instrument ID.
Note: Press GLP at any moment and the instruments will return to
measurement mode.
If calibration has not been performed, the instruments display
no CAL message blinking.
SETUP