beautypg.com

Hanna Instruments HI 2212 User Manual

Page 26

background image

26

CLEANING PROCEDURE

• General

Soak in Hanna HI 7061 or HI 8061 General
Cleaning Solution for approximately ½ hour.

• Protein

Soak in Hanna HI 7073 or HI 8073 Protein
Cleaning Solution for 15 minutes.

• Inorganic

Soak in Hanna HI 7074 Inorganic Cleaning
Solution for 15 minutes.

• Oil/grease

Rinse with Hanna HI 7077 or HI 8077 Oil
and Fat Cleaning Solution.

IMPORTANT:

After performing any of the cleaning procedures, rinse the

electrode thoroughly with distilled water, refill the reference chamber
with fresh electrolyte (not necessary for gel-filled electrodes) and soak the
electrode in HI 70300 or HI 80300 Storage Solution for at least 1 hour
before taking measurements.

This manual is related to the following products: