PRELIMINARY EXAMINATION

Remove the instrument from the packing material and examine it carefully to

make sure that no damage has occurred during shipping. If there is any

damage, notify your Dealer or the nearest Hanna Customer Service Center.

Each instrument is supplied with:

• HI 1131B Glass-body Combination pH Electrode with 1 m (3.3') Cable

• HI 7669/2W Temperature Probe

• HI 76404 Electrode Holder

• pH 4.01 & 7.01 Buffer Solutions (20 mL each)

• HI 7071S Electrolyte Solution

• 12VDC Power Adapter

• Instruction Manual

Note: Save all packing material until you are sure that the instrument

functions correctly. All defective items must be returned in the original

packing with the supplied accessories.

The HANNA HI 254 is a logging microprocessor-based pH and temperature

bench meter.

HI 254 can also measure Oxidation Reduction Potential (ORP) in mV range.

Relative mV feature is also provided.

pH measurements are compensated for temperature effect manually or

automatically with the HI 7669/2W temperature probe.

Up to five-point calibration is possible, using seven memorized buffers (1.68,

4.01, 6.86, 7.01, 9.18, 10.01 and 12.45 pH).

The GLP feature provides a guarantee of data consistency.

An alarm time-out is available to alert the user that too much time elapsed

since the last pH calibration.

The instrument can store data in memory at the user’s request for later

retrieval.

Hanna Instruments reserves the right to modify the design, construction

and appearance of its products without advance notice.