Hanna Instruments HI 254 User Manual
Page 19

19
BEEP STATUS SET
Press CAL when the beep status is displayed. Beep status (On or OFF) will
start blinking.
Press the ARROWkeys to change the beep status (On or OFF).
Press CFM to save the modified beep status.
Press CAL to escape without saving.
BAUD RATE SET
Press CAL when the baud rate is displayed. The baud rate (600, 1200, 2400,
4800 or 9600) will start blinking.
Press the ARROWkeys to change the baud rate value.
Press CFM to save the modified baud rate value.
Press CAL to escape without saving.
SERIAL COMMUNICATION COMMAND PREFIX SET
Press CAL when the command prefix is displayed. The command prefix (0 to
47) will start blinking.
Press the ARROWkeys to change the command prefix.
Press CFM to save the modified command prefix value.
Press CAL to escape without saving.
Note: See the PC interface section on page 25 for a complete explanation.
The command prefix must be 16 if HI 92000 PC software is used.
CALIBRATION ALARM TIME-OUT SET
Press CAL when the calibration time-out is displayed. Calibration time-out
(OFF or 1 to 7 days) will start blinking.
CAL
INTV