The Hanna HI 4011 and HI 4111 are ion selective elec-
trodes designed for the measurement of Iodide ions in
aqueous solutions. The HI 4011 is a solid state half-cell
sensor that requires a separate reference. The HI 4111 is a
combination ion selective electrode.

II

IIII

IIII.

Specifications

Specifications

Specifications

Specifications

Specifications

Type:

Solid State electrode with
a Silver Iodide pellet.

Ion(s) measured:

Iodide (I

- - - - -

)

Measurement range: 1.0 M to 1 X 10

-7

M

127,000 to 0.01 ppm

Interfering ions:

Cyanide and Mercury
must be absent. Strong
reducing solutions will
destroy membrane. Ratio
of interfering ion to I

-----

must

be less than the ratio
indicated below:

`

500 for Br

-----

Bromide

500 for Cl

-----

Chloride

Operating Temperature:

0-80

°C

Operating pH:

2.0-13.0 pH

Dimensions:

12 mm (OD) X 120 mm
nominal insertion
(0.47

” X 4.72”)

Connection:

BNC