Hi 4013 nitrate half-cell – Hanna Instruments HI 4013 User Manual
Page 2
![background image](/manuals/737264/2/background.png)
2
HI 4013 Nitrate Half-cell
HI 4013 Nitrate Half-cell
HI 4013 Nitrate Half-cell
HI 4013 Nitrate Half-cell
HI 4013 Nitrate Half-cell
HI 4113 Nitrate Combination Electrode
HI 4113 Nitrate Combination Electrode
HI 4113 Nitrate Combination Electrode
HI 4113 Nitrate Combination Electrode
HI 4113 Nitrate Combination Electrode
I.
I.
I.
I.
I. Introduction:
Introduction:
Introduction:
Introduction:
Introduction:
The Hanna HI 4013 and HI 4113 are ion selective elec-
trodes designed for the measurement of nitrate ions in
aqueous solutions. They utilize a replaceable sensing mod-
ule that contains an organic polymer membrane that is
sensitive to nitrate ions. The HI 4013 is a half-cell electrode
that requires a separate reference. The HI 4113 is a combi-
nation ion selective electrode.
II
IIII
IIII. Specifications
Specifications
Specifications
Specifications
Specifications
Type:
PVC membrane with or-
ganic ion exchanger
Ion measured:
Nitrate (NO
3
-
)
Measurement range:
0.1 M to 1X 10
-5
M
6200 to 0.62 ppm
Interference:
Organic solvents and cationic detergents must be absent.
Ratio of interfering ion to NO
3
-
must be less than the ratio
indicated below:
300
for F
-
fluoride
100
for Cl
-
chloride
4
for CO
3
2-
carbonate
2
for NO
2
-
nitrite
0.01
for I
-
iodide
0.0045 for ClO
4
-
perchlorate
Operating Temperature:
0-40°C
Operating pH:
3 to 8 pH (see Section
XIII)
Dimensions:
12 mm (OD) X 120 mm
nominal insertion
(0.47” X 4.72”)
Connection:
BNC