Procedure:
Follow sections VIII and IX to prepare electrodes for mea-
surement.
1) Follow section VI to prepare standards/ solution. Stan-

dards should bracket and fall within the range of
interest. Standards and solutions should be at the
same temperature.

2 mL of HI 4013-00 is added to 100 mL of both

samples and standards. OR

10 mL of ISISA are added to 50 mL both samples

or standards.
Add stir bar and mix before taking measurements.

2) Follow section VII; General Guidelines to optimize test

set-up.

3) During calibration it is best to start with lower concen-

tration samples first. Wait for a stable reading before
reading/ recording values. Permit longer equilibra-
tion times at these levels (3 or 4 minutes).

To prevent carry over and contamination of samples, rinse
sensors with deionized water and remove moisture with
absorbant tissue between samples.