Potassium HR

POTASSIUM HIGH RANGE

SPECIFICATIONS

Range

20 to 200 mg/L

Resolution

5 mg/L

Accuracy

±30 mg/L ±7% of reading at 25 °C

Typical EMC

±5 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 610 nm

Method

Adaptation of the Turbidimetric Tetraphenylborate method. The reaction between
Potassium and reagents causes turbidity in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93750

A-0

Potassium Reagent

6 drops

HI 93750

B-0

Powder Reagent

1 packet

REAGENT SETS

HI 93750-01 Reagents for 100 tests
HI 93750-03 Reagents for 300 tests
For other accessories see page 124.

MEASUREMENT PROCEDURE

Note: for sample preparation follow the CONCENTRATED SAMPLES procedure at page 18.

• Select the

Potassium HR method using the procedure

described in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of sample, up to the mark.

• Add six drops of HI 93750A-0, replace the cap and swirl

the solution.

• Place the cuvette into the holder and close the lid.

Phosphorus

• Add the content of one packet of HI 93706B-0 Phosphorus

Reagent B (Amino Acid) to the cuvette. Replace the cap
and shake gently until completely dissolved.

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 5 minutes and press

Read. When the timer ends the meter will perform the reading. The instrument

displays the results in

mg/L of phosphorus (P).

• Press ▲ or ▼ to access the second level functions.
• Press the

Chem Frm key to convert the result in mg/L of phosphate (PO

4

3

¯) and phosphorus pentoxide

(P

2

O

5

).

• Press ▲ or ▼ to return to the measurement screen.

INTERFERENCES

Interference may be caused by:
Sulfide
Chloride above 150000 mg/L
Calcium above 10000 mg/L as CaCO

3

Magnesium above 40000 mg/L as CaCO

3

Ferrous iron above 100 mg/L

10 mL