SPECIFICATIONS

Range

0.0 to 20.0 mg/L

Resolution

0.1 mg/L

Accuracy

±0.2 mg/L ±3% of reading at 25 °C

Typical EMC

±0.1 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, Periodate method. The reaction between manganese and reagents causes

a pink tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93709

A-0

Citrate

1 packet

HI 93709

B-0

Sodium periodate

1 packet

REAGENT SETS

HI 93709-01 Reagents for 100 tests
HI 93709-03 Reagents for 300 tests
For other accessories see page 124.

MEASUREMENT PROCEDURE

• Select the

Manganese HR method using the procedure described in

the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to the

mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the

Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cuvette.

• Add the content of one packet of HI 93709A-0 Citrate reagent.

Replace the cap and invert to mix with gently movements for
2 minutes.

Manganese HR

MANGANESE HIGH RANGE

• Remove the cuvette.

• Using the other 1 mL syringe, add to the cuvette exactly 0.5 mL of sample.

Note: Do not mix up the two syringes!

• Replace the cap and invert several times to mix.

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 15 seconds and press

Read. When the timer ends the meter will perform the reading.

• The instrument displays the results in

mg/L of Magnesium (Mg

2+

).

INTERFERENCES

Interferences may be caused by:

Acidity (as CaCO

3

) above 1000 mg/L

Alkalinity (as CaCO

3

) above 1000 mg/L

Calcium (Ca

2+

) above 200 mg/L

Iron must be absent

Aluminum must be absent

Copper must be absent

Magnesium

0.5 mL

of sample

10 mL